Open Access

Table 1

Instrument settings and furnace programs for analysis of Cu, Pb, As, Co, Ni, and Fe by GFAAS.

Conditions Cu Pb As Co Ni Fe
Wavelength (nm) 324.8 283.3 193.7 240.7 232 248.3
Lamp current (mA) 15 20 10 20 25 30
Air flow (ml min−1) 250 250 250 250 250 250
Injection volume (µL) 10 10 10 10 10 10
  Heating program temperature °C [ramp time (s), hold time (s)]
Drying 1 110 (2, 20) 100 (20, 10) 95 (20, 10) 110 (20, 0) 110 (20, 0) 110 (1, 30)
Drying 2 130 (5, 20) 190 (20, 10) 110 (20, 10) 250 (10, 0) 250 (10, 0) 130 (15, 30)
Pyrolysis 1 900 (10, 0) 900 (30, 20) 750 (20, 10) 1000 (10, 0) 1000 (10, 0) 1200(10, 0)
Pyrolysis 2 1200 (0, 13) 1200 (0, 15) 850 (0, 15) 1300 (0, 13) 1300 (0, 13) 1400 (0, 15)
Atomization 2000 (0, 2) 1600 (0, 5) 2450 (0, 5) 2300 (0, 2) 2150 (0, 2) 2400 (0, 5)
Cleaning 2200 (0, 2) 2400 (1, 3) 2500 (0, 2) 2400 (0, 2) 2250 (0, 2) 2500 (5, 2)

Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.

Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.